ClearanceBrief
Recall Z-0285-04 ยท IZH System, X-Ray, Mammographic

General Electric Med Systems

Seneographe 800T Mammography X-Ray Systems

DateJul 20, 2004 (initiated Feb 6, 2004)
StatusTerminated
ReasonThe mammography system's information for users lacked the technical specification for the minimum filtration permanently installed in the useful beam and the technical specification for the maximum line current.
Root causeOther
ActionThe consignees were sent 'Regulatory Notice' letters dated February 6, 2004. The letters informed the consignees of the violations and provided them with the technical specifications which needed to be provided.
DistributionNationwide in the United States
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953250SENOGRAPHE PEARL AND SENOGRAPHE SAPPHIRE
GE Medical Systems, Jan 26, 1996