Recall Z-0739-04 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Asahi Medical Co Ltd

Asahi APS Series Dialyzers, Model Nos. APS-400S, APS-550S, APS-650S, APS-750S, APS-900S, APS-1050S; the dialyzers are intended for single or initial use and under reprocessing and reuse conditions fo rup to 15 reuse cycles; Manufacturer: Asahi Medical Co., Ltd., 9-1, Kanda Mitoshirocho, Chiyoda-ku, Tokyo 101-8482, Japan; Distributed in teh USA by Nextron Medical, 45 Kulick Road, Fairfield, NJ 0700

DateJul 20, 2004 (initiated Oct 1, 2001)
StatusTerminated
ReasonCustomer reprocessing methods for the reuseable hollow fiber dialyzers may damage the hollow fibers and result in blood leaks during dialysis.
Root causeOther
ActionAsahi visited each of the U.S. customers starting on 10/1/01, observing how the facility handles the dialyzer before pre-cleaning, during pre-cleaning and priming. The users were retrained in the reprocessing steps, emphasizing the controls and care to be taken during pre-cleaning, including the verification of water pressure.
Quantity180,000 units estimated
DistributionNew Jersey, Ohio, Texas, California, Georgia, Pennsylvania, Illinois and Indiana
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001250APS SERIES DIALYZERS (WET MODELS), APS SERIES DIALYSIS (DRY MODELS)
Asahi Medical Co., Ltd., Aug 16, 2000