Recall Z-0738-04 ยท DTJ Holder, Heart-Valve, Prosthesis

Carbomedics, Inc

Carbomedics VT-200 Valve Tester

DateJul 20, 2004 (initiated Apr 30, 2002)
StatusTerminated
ReasonExcessive cleaning and resterilization cause cracking and crazing of device causing the device to malfunction during use.
Root causeOther
ActionLetters were sent to surgeons and hospital operating room supervisors in mid-May 2002 which included the following information: It is imperative that hospitals regularly and thoroughly inspect all testing instruments for indications of wear and degredation. And, CarboMedics is in the process of re-issuing suggested maintenance and sterilization guidlines.
Quantity4606 units
DistributionUS nationwide and to the following foreign countries: Denmark, Ireland, China, Cyprus, Greece, France, UK, Protugal, Ecuador, India, Japan, Hong Kong, Italy, Sweden, Belgium, Norway, Argentina, Spain, Malta, Austria, Switzerland, Netherlands, Germany, Australia, Finland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K960169CARBOMEDICS VALVE TESTER
Carbomedics, Inc., Apr 15, 1996