Recall Z-0589-04 ยท KZF Device, Medical Examination, Ac Powered

Burton Medical Products Corp

All Outpatient and Coolspot single ceiling or single track Medical Examination Lights manufactured before August 2000. Products manufactured during or after August 2000 contain a redesigned pivot assembly and are not included in this recall. Affected products include: Model 100540 Fleximount SGL Ceiling, Serial No.''s less than SC995486 Model 100740 SGL Ceiling CS II, Serial No''s less than SCCS

DateJul 20, 2004 (initiated Feb 19, 2004)
StatusTerminated
ReasonPivot joint on these ceiling mount lights may fail and lights may fall onto patients.
Root causeComponent design/selection
ActionFirm will ask for subdistribution information from thier distributors via letter on 1/30/2004. Letters will be sent directly to users. Firm will retrofit all sites with suspect units with new pivot assemblies.
Quantity17,000 estimated
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K851742MEDICAL EXAMINATION LIGHTS
Burton Medical Products Corp., May 13, 1985