Recall Z-0549-04 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Becton Dickinson & Company

VACUTAINER Brand Push Button Blood Collection Sets.

DateFeb 25, 2004 (initiated Dec 17, 2003)
StatusTerminated
ReasonA hole was noticed in the 'blister pack' that is used to package the device and provides the sterility barrier. Sterility could be compromised.
Root causeOther
ActionBD Sales Organization directly contacted all customers with return instructions on December 22, 2003. All consignees responded.
Quantity4400
DistributionThe firm was able to determine that the product was distributed to 3 customers: Long Beach Memorial, Long Beach, CA; Potomac Hospital, Woodbridge Prince William, VA; and St. Joseph Hospital, Orange, CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011984BD VACUTAINER PUSH BUTTON BLOOD COLECTION SET
Becton, Dickinson & CO, Aug 29, 2001