ClearanceBrief
Recall Z-0383-04 ยท DZI Drill, Bone, Powered

Implant Innovations, Inc.

Products are various dental surgical drills packed in a sealed tray and labeled as sterile. The tray is labled as follows: 3i Implant Innovations, Inc., 4555 Riverside Drive, Palm Beach 33410 USA 561-776-6700 DC500 LOT 5.0mm Implant Disposable Countersink Pilot Drill STERILE

DateJul 20, 2004 (initiated Dec 2, 2003)
StatusTerminated
ReasonThe tray lid seals may be open thereby compromising the sterility barriers for the recalled lots.
Root causeOther
ActionThe firm notified international distributors on 12/2/2003 via e-mail and also a recall notification letter was faxed to them on 12/2/2003. They were instructed to subrecall to the health professional level. Also on 12/2/2003 all domestic accounts (health practioners) were telephoned and then faxed a recall notification letter. Each account was instructed to check their inventory, respond back by fax and to return recalled product to Implant Innovations Inc.
Quantity106
DistributionProduct was distributed to 45 Dentist/Surgeons nationwide in the US and also to 13 international Distributors in Canada, Sweden, Germany, Colombia, Spain, Switzerland, Australia, Japan, Austria, France, Poland, England and Greece.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.