Recall Z-0517-04 ยท BTM Ventilator, Emergency, Manual (Resuscitator)

Laerdal Medical Corporation

LAERDAL Silicone Resuscitator (LSR), Adult size, containing Intake/Reservoir Valve. Sold as Part Nos: 870050-Adult Basic w/o Mask in Carton; 870051-Adult Complete in Carton; 870052-Adult Standard with Adult Mask 4-5+ in Carton; 870053-Adult Complete in Compact Case; 870055-Adult Complete in Display Case; 875500-Adult Update Kit. Responsible firm on the label: Distributed in the U.S. by Laerdal

DateFeb 10, 2004 (initiated Dec 23, 2003)
StatusTerminated
ReasonPotential for the Flap Valve component of the Intake/Reservoir Valve to become dislodged from its mount, rendering the resuscitator inoperative.
Root causeOther
ActionLetters issued on 1/14/2004. The recall involves the replacement of the Inner Part of the Intake/Reservoir Valve, which is part of the Laerdal Silicone Resuscitator. Recall is to the user level.
Quantity2,153 units
DistributionProduct was shipped to approximately 264 customers nationwide, and 3 foreign consignees (Mexico, Brazil and Northern Mariana Islands). Product was also sold to a few exporters who shipped units to unknown countries. Customers include distributors and end users. There were 2,403 parts shipped to