ClearanceBrief
Recall Z-0348-04 ยท NOM Antigen, Galactomannan, Aspergillus Spp.

Bio Rad Laboratories, Inc.

Platelia Aspergillus EIA

DateJan 14, 2004 (initiated Nov 6, 2003)
StatusTerminated
ReasonPotential drug and device interaction with piperacillin/tazobactam (Zosyn) causing positive test result for galactomannan.
Root causeOther
ActionCustomers were sent a letter, Technical Bulletin and revised package insert on November 6, 2003.
Quantity147 kits
DistributionThe firm distributed to hospitals and diagnostic laboratories nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023857PLATELIA ASPERGILLUS EIA, MODELS 62793 AND 62794
Bio-Rad, May 16, 2003