Recall Z-0538-04 ยท DSI Detector And Alarm, Arrhythmia

Welch Allyn Protocol, Inc

Welch Allyn Acuity Central Monitoring Station (With SunBlade Model 150,650 MHZ. Part # 700-0362-00

DateFeb 13, 2004 (initiated Dec 31, 2003)
StatusTerminated
ReasonPotential failure to audibly alarm due to speaker hardware component failure.
Root causeOther
ActionThe firm sent a letter dated December 31, 2003 instructing their customers to install an audio cable from the CPU to the Flat Panel Monitor to provide redundant speaker capability. The audio cable and installation instructions were included in the letter. The letter also included a Correction Response Form which customers should fill out upon completion of correction.
Quantity233 devices
DistributionThe firm distributes the product nationwide and internationally.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022453MODIFICATION TO ACUITY CENTRAL MONITORING SYSTEM
Welch Allyn Protocol, Inc., Aug 21, 2002