Recall Z-0353-04 ยท CEC Radioimmunoassay, Free Thyroxine

Abbott Laboratories HPD/ADD/GPRD

ARCHITECT Free T4 Calibrator Kit, list 7G96-01; the kit consists of 2 bottles of calibrators, CAL 1 - 0.5 and CAL 2 - 6.0; Abbott Laboratories, Abbott Park, IL 60064

DateJul 20, 2004 (initiated Dec 15, 2003)
StatusTerminated
ReasonThe calibrator lots may yield Free T4 values that are depressed.
Root causeOther
ActionArchitect Free T4 Calibrator kits were recalled by letter dated 12/15/03. The accounts were informed of the out of range control values, and the potential for depressed patient values with the kits, and were requested to discontinue use and destroy the affected lots of product.
Quantity1,364 kits
DistributionNationwide and internationally to Abbott subsidiaries in Brazil, Canada, Germany, Japan, Singapore, Hong Kong, Thailand, Korea, Australia, Colombia, New Zealand, Venezuela and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983417ABBOTT ARCHITECT FREE T4
Abbott Laboratories, Nov 20, 1998