Recall Z-0298-04 ยท JHX Fluorometric Method, Cpk Or Isoenzymes

Ortho-Clinical Diagnostics

Vitros Immunodiagnostic Products CK-MB REAGENT PACK, REF 189 6836, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.

DateJan 6, 2004 (initiated Nov 26, 2003)
StatusTerminated
ReasonReagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
Root causeOther
ActionCustomers notified via phone and faxed letter on 11/26/03. Customers provided with a modified QC procedure to be used until new shipments can be made.
Quantity774 reagent packs
Distribution738 reagent packs shipped to U.S. consignees and 36 reagent packs to Canadian consignees.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993068VITROS IMMUNODIAGNOSTIC PRODUCTS CK-MB REAGENT PACK ; VITROS IMMUNODIAGNOSTIC PRODUCTS CK-MB CALIBRATORS