Recall Z-0255-04 ยท DYB Introducer, Catheter

Arrow International Inc

Arrow A Port with Attachable Silicone Rubber Catheters Implantable Vascular Access System (Product # AP-01510) . Product # AP-01510 kit contains 1 implantable titanium port, 2 attachable silicone rubber catheters (8.4 Fr. x 75 cm and 9.6 Fr x 75 cm) straight non-coring needle, patient ID card and chart sticker. Kit sold individually or per case (5 trays/case).

DateJul 20, 2004 (initiated Nov 4, 2003)
StatusTerminated
ReasonThe 9.6 Fr. x 75 cm silicone rubber catheter was a higher durometer (less flexible) than previously provided.
Root causeOther
ActionThe firm began notification of consignees via U.S. mail on 11/4/03. A notice of the recall was also posted on the firm''s website (search for 'recall').
Quantity913
DistributionProduct was distributed internationally (France, Germany, Mexico, United Kingdom, Singapore, Brazil, Australia) and domestically. Domestic distribution was concentrated in the Midwest and Eastern portions of the US. Total direct accounts : Distributors/dealers, international - 7 Distributors/d