Recall Z-0240-04 ยท CDQ Urease And Glutamic Dehydrogenase, Urea Nitrogen

Abbott Laboratories Diagnostic Div

AEROSET/ARCHITECT c8000 Urea Nitrogen Models: 7D75-01; 7D75-20; 7D75-30

DateDec 18, 2003 (initiated Nov 24, 2003)
StatusTerminated
ReasonFalse low results may occur with out error codes due to substrate depletion in samples with more than 220mg/dL BUN.
Root causeOther
ActionFirm sent letter to customers dated November 24,2003 which includes a new package insert to use with this reagent. Instructions were changed to correct low resulting situation. Subrecall was requested.
QuantityAll lots.
DistributionNationwide and Mexico, Brazil, Germany, Singapore, Grand Cayman, Venezuela, Puerto Rico, Hong Kong, Australia, Columbia, Dominican Republic, China, Trinidad, Argentina, Canada, New Zealand, Bahamas