Recall Z-0229-04 ยท LPO Gases Used Within Eye To Place Pressure On Detached Retina

Alcon Laboratories, Inc

ISPAN patient bracelet, given to patients post surgery to warn of further use of dinitrogen monoxide

DateDec 17, 2003 (initiated Nov 18, 2003)
StatusTerminated
ReasonThe firm initiated the recall on 11/18/2003 via letters to all consignees. The letter advises physicians to stop using any patient bracelets (blue color) that contain the word 'stickoxid', destroy all blue bracelets, and identify all patients who had either ISPAN gas administered who may still have a bubble present in the eye.
Root causeOther
ActionThe patient bracelet that is given to patients post surgery after administering either ISPAN Perfluoropropane or ISPAN Sulfur Hexafluordie referred to the term, 'stickoxid', which is the general term for nitrogen oxide and may be interpreted as either nitrogen monoxide (NO) or dinitrogen monxoide. The firm is recalling the old (blue) bracelets and replacing them with new (grey) bracelets that now reference the specific term dinitrogen monoxide N20 (laughing gas).
DistributionGermany, Switzerland, and Austria
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.