Recall Z-0263-04 ยท JQP Calculator/Data Processing Module, For Clinical Use

Abbott Laboratories, Inc

The ARCHITECT c8000, Clinical Chemistry Analyzer, Manufactured by Abbott Laboratories, Irving, Texas

DateDec 24, 2003 (initiated Nov 17, 2003)
StatusTerminated
ReasonSample Identification (SID) numbers of a bar coded sample may be truncated, or shortened, after being scanned by the bar code reader (e.g. an actual SID of 123456 is shortened to 1234)
Root causeOther
ActionThe firm plans to distribute a Product Information Letter and an updated ARCHITECT i System Operations Manufal Addendum to all US ARCHITECT customers. An electronic version of these documents was sent to all Abbott affilities worldwide for distribution to their ARCHITECT customers.
Quantity148
DistributionMexico, Puerto Rico, Japan, Taiwan, Australia, Colombia, Cananda, Hong Kong, New Zealand, Netherlands, Singapore, Germany, Grand Cayman, Korea, Chile, Thailand, TX, PA, SC, CA, IN.
FDA recordOpen on fda.gov