Recall Z-0289-04 ยท DQA Oximeter

Minolta Corp

Pulsox-3Si. LCD, SpO2, PR, Pulse strength bar, Motion artifact, Low battery message, Error message, Event mark, Measuring range Sp02 (%) 0 to 100, Pulse rate 20 to 250, Measuring accuracy +/- 2% (70 to 100), Pulse rate +/- 2%, Sp02 alarm, no, Probe type, finger, multi-site, Patient range adult to neonatal, memory function Approx. 24 hrs, printer/recorder optional accessory, output RS232 w/optiona

DateJul 20, 2004 (initiated Nov 14, 2003)
StatusTerminated
ReasonError in the internal software. When a pulse level is high or subject has SPO2 of less than 80%, the data shifts.
Root causeOther
ActionDistributor provided customer list to Konica Minolta who initiated phone calls to the customers on 11/14/2003.
Quantity291
DistributionThe units were shipped to one distributor in Utah.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010413MINOLTA PULSOX-3SI, PULSOX-3IA, PULSOX-3LI
Konica Minolta Sensing, Inc., Aug 2, 2001