Recall Z-0143-04 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Datex-Ohmeda Inc
Datex-Ohmeda Cardiocap/5 Patient Monitors
| Date | Jul 20, 2004 (initiated Oct 17, 2003) |
| Status | Terminated |
| Reason | Datex-Ohmeda Cardiocap/5 Patient Monitors have bracket pins which do not adequately secure them to their optional mounting stand so the monitors may fall. |
| Root cause | Other |
| Action | Recall letters dated October 15, 2003 were sent to the consignees on 10/17/03. The letters request that the consignees order service kits to correct the units and use the kits to make the corrections of the devices. |
| Quantity | 875 units |
| Distribution | The monitors were shipped to consignees located nationwide in the United States and to foreign consignees located worldwide. |
| FDA record | Open on fda.gov |