Recall Z-0092-04 ยท DSI Detector And Alarm, Arrhythmia

General Electric Medical Systems Information Technology

Solar 8000M Patient Monitor with software versions 3A, 3B and 3C

DateNov 6, 2003 (initiated Jun 14, 2002)
StatusTerminated
ReasonVarious user settings may change without an audible warning or visual indication. Examples of settings that may change are pacemaker detection which may turn from on to off, selected ECG leads for display, and selected gain settings for a selected waveform.
Root causeOther
ActionA 'Patient Safety Alert' letter dated June 14, 2002 was sent to the consignees. The letter recommended certain actions to avoid the problems and stated that the firm''s field service team would contact the consignees to schedule a software revision to eliminate the problem.
Quantity2809 monitors and 238 software update kits
DistributionThe devices were distributed nationwide in the United States and worldwide.