Recall Z-0070-04 ยท LFL Instrument, Ultrasonic Surgical

Misonix, Inc

LySonix 2000 Ultrasonic Surgical Systems, MODEL NO. LY-2000. --- Old labeling: MANUFACTURED BY: MISONIX, INC. EXCLUSIVELY FOR SALE BY MENTOR CORPORATION UNDER EXCLUSIVE LICENSE FROM SONIQUE SURGICAL, INC. MISONIX, INC. 1938 NEW HIGHWAY, FARMINGDALE, NY 11735. Device was manufactured from September 1996 to present.

DateOct 30, 2003 (initiated Aug 13, 2003)
StatusTerminated
ReasonFirm received complaints of tingling or static shock sensation to end users caused by forcing the output connector into its receptacle misaligned. The misalignment causes additional voltage to appear on the patient probe.
Root causeOther
ActionMisonix, Inc. notified the sole direct consignee, Byron Medical Inc., of the "Corrective Action" by letter on 8/13/03. On 9/17/03, Byron Medical Inc. notified the end users by letter, and provided instructions for proper assembly and warning labels to be affixed to the units.
Quantity1119 units (potentially)
Distribution1 distributor in Tucson, AZ, for further distribution to end users (physicians and medical facilities) nationwide & Puerto Rico, and in the following countries: Canada, Mexico, Australia, Switzerland, Israel, Saudi Arabia, Cyprus, Greece, Turkey, Germany, Dominican Republic, Sweden, Netherlands, Chi