Recall Z-0288-04 ยท FMZ Incubator, Neonatal

Ohmeda Medical, A Div of Datex-Ohmeda , Inc

Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed.

DateJul 20, 2004 (initiated Oct 17, 2003)
StatusTerminated
ReasonHeater vent door on neonatal incubator may malfunction.
Root causeOther
ActionThe recalling firm notified domestic consignees by certified letter and international distributors by e-mail dated 10/17/2003. The recall notice was incorporated into a firm''s follow-up letter concerning an on-going product correction of the elevating base on the same neonatal beds. The letter advises of the malfunctioning heater door vent and states that the firm will replace all existing vent doors with doors made of higher temperature rated material. Firm will provide retrofit kits for consignee or service representative installation.
Quantity1404 beds
DistributionProducts were sold to 365 direct accounts around the US and 58 international distributors.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020543OHMEDA MEDICAL GIRAFFE OMNIBED
Ohmeda Medical, May 9, 2002