Recall Z-0099-04 ยท EHD Unit, X-Ray, Extraoral With Timer

Philips Medical Systems Sales & Service Region No. America

Clea Stand Ceiling used with the Integris X-Ray System H5000, Allura 12/15, Allura 9C

DateJul 20, 2004 (initiated Oct 10, 2003)
StatusTerminated
ReasonThe cable carrier and cables may fall after repeat collisions with OR lamps, radiation shields or other supported accessories.
Root causeOther
ActionThe firm sent a letter dated, October 10, 2003 indicating the firm''s service representatives will upgrade the cable carrier by mounting a safety catch inside the cable carrier.
Quantity536 units
DistributionThe firm distributes to hospitals and medical centers nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K971365PHILIPS INTEGRIS H5000
Philips Medical Systems North America, Inc., May 14, 1997