Recall Z-0097-04 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems Sales & Service Region No. America

Monitor Ceiling Suspension for Fluoroscopic and Angiographic X-Ray Systems.

DateJul 20, 2004 (initiated Oct 10, 2003)
StatusTerminated
ReasonThe thumbscrew hinge pin on the monitor support arm could become loose allowing the monitor to fall.
Root causeOther
ActionThe firm issued a letter dated October 10, 2003 to all affected customers. The suspension will be upgraded by field service personnel starting October 24, 2003.
Quantity570 units
DistributionThe firm distributes to hospitals and medical centers.