Recall Z-0064-04 ยท DXT Injector And Syringe, Angiographic

Acist Medical Systems

ACIST Automated Manifold (Transducer) Kit, Model BT2000, SKU #014613 (used with ACIST Contrast Management Systems (CMS2000) and E2000 Voyager Units)

DateOct 24, 2003 (initiated Oct 2, 2003)
StatusTerminated
ReasonSome of the automated manifold's 2.5 inch high-pressure lines for radiopaque contrast media, used in angiographic procedures, have separated where they are bonded to the manifold body of the device allowing contrast media to leak.
Root causeOther
ActionThe consignees were sent letters dated 10/2/03 which requested that the product be returned.
Quantity7120 kits
DistributionNationwide in the United States.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K984231ANGIOGRAPHIC CONTRAST MANAGEMENT SYSTEM, MODEL PART A AND PART B
Invasatec, May 31, 2000
K991103MODIFICATION TO ACIST CL 100H
Acist Medical Systems, May 31, 2000