Recall Z-0063-04 ยท MLM Enzyme Immunoassay, Tracrolimus

Abbott Laboratories HPD/ADD/GPRD

IMx Tacrolimus II Calibrators, list 3C10-10; the pack contains six 4.5-mL bottles, with Calibrator A having 0 ng/mL, Calibrator B having 3 ng/mL, Calibrator C having 6 ng/mL, Calibrator D having 12 ng/mL, Calibrator E having 20 ng/mL and Calibrator F having 30 U/mL; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park, IL 60064

DateOct 24, 2003 (initiated Sep 19, 2003)
StatusTerminated
ReasonInvalid test results are being generated due to calibration errors when using calibrator lots 94165M100 and 94164M100.
Root causeOther
ActionRecalled by letters dated 9/19/03 and 10/7/03. The accounts were informed of the inability to calibrate the IMx Tacrolimus II assay when using calibrator lots 94165M100 and 94164M100, and initially requested those accounts experiencing an increase in calibration-related errors to report the issue to Abbott for a replacement calibrator kit. The 10/7/03 letter requested the accounts to discontinue use and destroy any of the two affected calibrator lots found in inventory.
Quantity2677 kits
DistributionNationwide and internationally to Guatemala, Venezuela, Peru, Argentina, Dominican Republic, Canada, Japan, Singapore, Thailand, Korea, Australia, Costa Rica, Panama and Great Britain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.