Recall Z-0106-04 ยท KPS System, Tomography, Computed, Emission

Cti Pet Systems Inc

Siemens Models: BIOGRAPH, BIOGRAPH LSO and CTI Models: REVEAL HD, REVEAL RT

DateNov 13, 2003 (initiated Oct 2, 2003)
StatusTerminated
ReasonCPS has found that in some instances, when the CT E-Stop is activated, a malfunction may occur where the PHS controller may not shut down and stop bed motion.
Root causeOther
ActionWritten notification was sent out beginning 10/02/2003 to distributors and end users to advise them of the problem.
Quantity73
DistributionNationwide International: Japan, Germany, Brazil, Spain, China, Russia, Austrailia, Hong Kong, Singapore, Poland France, Korea, Greece, UK and Austria