Recall Z-0060-04 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Nova Biomedical Corporation

Stat Profile Ultra Magnesium Sensor (In-vitro Diagnostic used with the Stat Profile Ultra M analyzer) Part # 20394

DateOct 24, 2003 (initiated Sep 22, 2003)
StatusTerminated
ReasonDegradation of the magnesium sensor could result in low ionozed magnesium patient samples
Root causeOther
ActionNova Biomedical contacted customers by telephone beginning on 9/26/03. Each account was faxed or mailed Customer Information Bulletin requesting the discard of Mg++ sensors in inventory that are greater than 4 months old from date of manufacture. Replacement product will be issued.
Quantity659 units
DistributionNatinwide Foreign: Italy, Czech, Japan, Taiwan, Hungary, India, Israel, Mexico, Canada, Germany,Switzerland, Tunisia, Korea, Brazil, Greece, Coatia, China, France, Turkey, Thailand, Argentina. Government: VA Decator, GA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981427STAT PROFILE M ANALYZER
Nova Biomedical Corp., May 12, 1998