Recall Z-0172-04 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas

DateJul 20, 2004 (initiated May 8, 2002)
StatusTerminated
ReasonIf the interlock cable connected to the door lid sensor is disconnected or fails, the lid is detected as closed even though it may be open.
Root causeOther
ActionFirm sent technical service bulletins with corrective instructions to consignees with all existing ARCHITECT i 2000 processing modules in February 2003.
Quantity1,167 units worldwide
DistributionStates of Virginia, Illinois, New York, Texas, Oklahoma, Pennsylvania, ILlinois, Michigan, California, New Hampshire, Ohio, Florida, Maryland, Nevada, Georgia, Minnesota, Colorado, Arizona, Missouri, Louisiana, Oregon, Wisconsin, Hawaii, Connecticut, New Hampshire, Massachusetts, Washington, South