Recall Z-0165-04 ยท DFT Igm, Antigen, Antiserum, Control

Abbott Laboratories Diagnostic Div

AEROSET Immunoglobulin M, list number 1E01-01, manufactured by Abbott Laboratories, South Pasadena, CA

DateJul 20, 2004 (initiated Jul 31, 2002)
StatusTerminated
ReasonAssay may produce falsely low IgM results without error codes when testing specimens contain elevated serum paraproteins associated with various myeloma or other malignancies.
Root causeOther
ActionConsignees were provided with assay parameters for the RCD flag which may enhance the detection of these type of specimens. The assay parameter settings for the RCD flag were distributed to all AEROSET consignees in the United States by direct mail and to country managers worldwide via a Product Information letter dated July 31, 2002. Also provided with the Product Information letter was an AEROSET IgM Application Sheet containing the updated assay parameters for the RCD flag.
Quantity205 units
DistributionNationwide to the following states: Texas, Illinois, Indiana, Tennessee, Alaska, Pennsylvania, New York, Florida, Massachusetts, Missouri, Alabama, North Carolina, California, Ohio, Utah, Michigan, Arizona, Wisconsin, Oklahoma, Maine, Kansas, Georgia, Mississippi, Connecticut, Rhode Island, Washing