Recall Z-0052-04 ยท DXY Implantable Pacemaker Pulse-Generator

St Jude Medical

Pulse Generator

DateOct 24, 2003 (initiated Jul 31, 2003)
StatusTerminated
ReasonUnder certain circumstances, pacemaker has a potential to deliver a short coupled pacing interval of approx. 300 msec (200ppm). It is also possible that a patient may experience (up to max of 12) shorter than anticipated pacing intervals.
Root causeOther
ActionSales Representatives will personally contact the physicians, explain the situation, leave a technical memo and a list of implanted patients for the doctor to follow-up with. The ''Dear Doctor'' letter to accompany the representatives is dated 7/31/2003. Instruments not implanted will be returned for reprocessing.
Quantity775
DistributionNationwide and to four VA Medical Centers in IA and MO.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.