Recall Z-1234-03 ยท LPN Accessories, Soft Lens Products

Ciba Vision Corporation

Quick Care Starting Solution. The product is used for cleaning and disinfection of soft (hydrophilic) contact lenses.

DateSep 13, 2003 (initiated Aug 13, 2003)
StatusTerminated
ReasonInvestigation has determined that the concentration of the isopropyl alcohol (IPA) decreased over time and may fall below specifications prior to reaching the expiration date.
Root causeOther
ActionThe consignees were notified of the recall by letter on August 13, 2003 asking for return of the product.
Quantity2,136 units
DistributionCA, CT, FL, GA, IA, KS, MI, MN, OH, and WI
FDA recordOpen on fda.gov

510(k) clearances named in this recall