Recall Z-1094-03 ยท LSL Dna-Reagents, Neisseria

Becton Dickinson & Co.

Becton Dickinson's ProbeTec (tm) ET Instrument

DateAug 20, 2003 (initiated Jul 21, 2003)
StatusTerminated
ReasonComponent of in vitro diagnostic device was incorrectly installed causing incorrect reporting of patient results.
Root causeOther
ActionThe recalling firm notified consignees by letter on 7/21/2003 and advised of the potential for incorrect optical bundle installation. The firm provided an optical test procedure with materials to conduct an on site performance test. Consignees were instructed not to report patient results until optical performance has been verified. Firm will replace or repair failing instruments.
Quantity1015 units
DistributionProduct was distributed to domestic and military end users nationwide and international distributors.