Recall Z-0051-04 ยท G-I

Propper Mfg Co Inc

propper brand cal-mark(R), Micro-Hematocrit Calibrated Capillary Tubes, CALIBRATED-HEPARINIZED, reorder no. 043021, 6 Vials, 100 Tubes per Vial, Length 75 mm., 0.55 mm. I.D., 1.40-1.75mm. O.D. Calibrated 60mm from end for use in direct reading centrifuges. Fill to the 60mm mark with blood and seal opposite end by flame or with sealing compound. If centrifuged on a direct reading centrifuge, re

DateOct 24, 2003 (initiated Aug 9, 2002)
StatusTerminated
ReasonDeviation in capillary tubes could cause a 10% increase in read out values which could make a slight anemic patient appear to have a normal reading.
Root causeOther
ActionOn 8/9/02, Propper faxed a 'Request for Return' of the product to the consignees advising them of the problem and that a replacement will be sent for all returned product. On 11/27/02, Propper faxed a second 'Request for Return of Product' to its consignees. The firm destroyed the recovered and held stock.
Quantity178 boxes
DistributionDistributors located in TX, UT, ID, IL, IA, IN, MN, WI, OH, CT, MO, PA, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.