ClearanceBrief
Recall Z-1124-03 ยท GEX Powered Laser Surgical Instrument

Intralase Corp

Intralase FS Laser System. Laser Keratome

DateAug 12, 2003 (initiated Jul 15, 2003)
StatusTerminated
ReasonGantry moves toward patient when 'Home' button is used after procedure.
Root causeOther
ActionFirm sent Safety Alert letters to customers on July 15, 2003 by registered mail instructing not to use the HOME button, including a 'Do Not Use' sticker for the machine and an addendum to the Operations Manual.
Quantity70
DistributionNationwide and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001211MODIFICATION TO INTRALASE 600C LASER KERATOME
Intra Lase Corp., Dec 4, 2000
K002890INTRALASE 600C LASER KERATOME
Intra Lase Corp., Aug 9, 2001
K013941PULSION FS LASER KERATOME
Intra Lase Corp., Feb 27, 2002
K993153INTRALASE 600C LASER KERATOME
Intra Lase Corp., Dec 17, 1999