Recall Z-1164-03 ยท FPA Set, Administration, Intravascular
Baxter Healthcare Corp.
Colleague Pump Syringe Adapter Set, 92'', Injection Site, Non-DEHP Tubing, Luer Lock Adapters, catalog 2C0169; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
| Date | Aug 21, 2003 (initiated Jul 16, 2003) |
| Status | Terminated |
| Reason | There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process. |
| Root cause | Other |
| Action | Recalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product. |
| Quantity | 12,576 units |
| Distribution | Nationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K982672 | COLLEAGUE PUMP SYRINGE ADAPTER SET Baxter Healthcare Corp, Sep 4, 1998 |