ClearanceBrief
Recall Z-1164-03 ยท FPA Set, Administration, Intravascular

Baxter Healthcare Corp.

Colleague Pump Syringe Adapter Set, 92'', Injection Site, Non-DEHP Tubing, Luer Lock Adapters, catalog 2C0169; Baxter Healthcare Corporation, Deerfield, IL 60015 USA

DateAug 21, 2003 (initiated Jul 16, 2003)
StatusTerminated
ReasonThere are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
Root causeOther
ActionRecalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product.
Quantity12,576 units
DistributionNationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982672COLLEAGUE PUMP SYRINGE ADAPTER SET
Baxter Healthcare Corp, Sep 4, 1998