Recall Z-1072-03 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Diametrics Medical, Inc.

IRMA SL Blood Analysis System CC pH, pCO2, pO2, Hct, Na+, K+, iCa Cartridge, Part Number 039903(M3586A). There are 25 disposable cartridges in a box of IRMA SL Blood Analysis System CC pH, pCO2, pO2, Hct, Na+, K+, iCa Cartridges, Part Number 039903(M3586A).

DateJul 30, 2003 (initiated Jul 14, 2003)
StatusTerminated
ReasonAn incorrect lot designation, when entered into the blood analysis device, caused the device to give an error message and not operate.
Root causeOther
ActionThe consignees were notified of the recall by telephone on July 14, 2003 and requested to return the product.
Quantity103 boxes of 25 cartridges each.
DistributionThe product was shipped to 13 hospitals in California, Colorado, Florida, Iowa, Kansas, Minnesota, New Hamshire, Texas, and Wisconsin.