Recall Z-1126-03 ยท LMB Device, Digital Image Storage, Radiological

Olympus America Inc.

Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4 --- Common name: Medical image storage device. Class I, 510 (k) exempt.

DateAug 12, 2003 (initiated Mar 5, 2003)
StatusTerminated
ReasonSoftware defect; potential under certain circumstances to incorrectly identify endoscopy images that have been uploaded & saved to the network system.
Root causeOther
ActionOlympus America, Inc. sent "Urgent-Device Correction" letters, dated 3/5/03, to the medical facilities, advising them of the software malfunction. The letter provides instructions to health care practitioners and that a new software release is being developed to prevent a recurrence.
Quantity856
DistributionMedical facilities nationwide and Puerto Rico; and 4 foreign accounts in Canada, Ecuador, Chile & Mexico.