ClearanceBrief
Recall Z-1014-03 ยท DXY Implantable Pacemaker Pulse-Generator

Guidant Corp-Cpi Division

Discovery II Multiprogrammable Pacemaker, Model 1283

DateJul 15, 2003 (initiated Jun 23, 2003)
StatusTerminated
ReasonSome of the pacemakers could have failed to pace due to an inadequately soldered electrical connection at an integrated circuit.
Root causeOther
ActionOn June 23, 2003, sales representatives began retrieving any recalled pacemakers, which were not implanted. For implanted pacemakers, an 'Important Patient Management Information' letter, dated June 23, 2003, was sent to physicians recommending that pacemaker-dependent patients, with the potentially affected pacemakers, be scheduled for exams to check for proper pacemaker performance using a programmer.
Quantity3 pacemakers
DistributionNationwide in the United States and worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.