Recall Z-1047-03 ยท LOK Kit, Test, Alpha-Fetoprotein For Neural Tube Defects

Abbott Laboratories HPD/ADD

AxSYM AFP Reagent Pack, list 7A48-22 (100 test kit), 7A48-21(French label); and AxSYM Cancer Assay Disk, list 3D50-02 (Version 2.0), 3D50-03 (Version 3.0), and 3D50-04 (Version 4.0)

DateJul 24, 2003 (initiated Jun 12, 2003)
StatusTerminated
ReasonAFP results obtained when using the 1:101 automated dilution protocol showed an overestimation of AFP of up to 30.5%
Root causeOther
ActionAll AxSYM AFP customers were informed not to use the 1:101 automated dilution protocol via Device Correction letter dated 6/12/03. If necessary, manually dilute samples per the AxSYM AFP package insert.
Quantity23,000 kits
DistributionNationwide and internationally to Australia, Canada, Costa Rica, Ecuador, Japan, Panama, San Salvador, Thailand, Barbados, Chili, Curacao, Honduras, Korea, Peru, Singapore, United Kingdom, Brazil, Columbia, Dominican Republic, Hong Kong, Pakistan, St. Vincent, Uruguay
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.