Recall Z-0989-03 ยท INX Bed, Air Fluidized

Hill-Rom Manufacturing, Inc.

CLINITRON CII Air Fluidized Therapy unit.

DateJul 10, 2003 (initiated May 5, 2003)
StatusTerminated
ReasonMalfunction of side-rail latching pin coupled with inadequate directions for use.
Root causeOther
ActionConsignees were notified by letter sent via Airborne express, on May 5, 2003. Side-rail latching pins will be replaced with a more robust version. Label indicating directions for use will be modified to include detailed information on proper procedures for moving the units.
Quantity122 units (including both units)
DistributionThe units are primarily leased to hospitals. Units have been sold to hospitals located nationwide and to Canada, Austria, France, Germany Ireland Italy, Netherlands, Spain, Sweden, Switzerland and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K813584CLINITRON AIR FLUIDIZED SUPPORT SYSTEM
Uhi Corp., Jan 12, 1982