Recall Z-0982-03 ยท BTM Ventilator, Emergency, Manual (Resuscitator)

Ventlab Corporation

V*CARE, Infant Resuscitator, with 40 cm HO Pressure Limiting System & Pressure Port Adapter, VN 3000 Series, Non-Sterile. The product included a clear plastic bag with white background which included Direction for Use statement #5. To remove vomitus: Disconnect resuscitator from patient. Tap the patient valve several times while squeezing the bag body. Retest the resuscitator for proper functionin

DateJul 8, 2003 (initiated Apr 15, 2003)
StatusTerminated
ReasonPrinted bag with instructions to verify the pressure with a 'certified manometer' was omitted in some of the product.
Root causeOther
ActionConsignee was notified by fax on 3/15/2003. All product on hand were returned to Ventlab.
Quantity9 cases (10 units per case)
DistributionThe product was distributed to one consignee in CA. The cases were returned to the recalling firm.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.