Recall Z-0652-03 ยท KPR System, X-Ray, Stationary

Sedecal USA, Inc.

Optima URS X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) SPAIN

DateApr 25, 2003 (initiated Sep 24, 2002)
StatusTerminated
ReasonIncorrect identification labels and their dertification testing program faialed in noncompliance with U.S. Federal performance standards.
Root causeOther
ActionSedecal USA sent end user notification letters dated 9/24/02 to the consignees having the affected Sedecal Stationary X-Ray Systems, informing them that the units do not meet with the U.S. federal performance standards, that the labels are in Spanish and that the labels may not be visible after system installation. Sedecal schedule an appointment to conduct performance tests on the mobile x-ray systems to verify their compliance with the federal performance standard, submit the data to FDA, and place the English labels on the units. Sedecal will reimburse the owners for any out-of-pocket exp
Quantity5 units
DistributionNew York, Texas, Florida and Minnesota.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012546OPTIMA URS UNIVERSAL RADIOGRAPHIC SYSTEM
Sedecal USA, Inc., Nov 2, 2001