ClearanceBrief
Recall Z-0864-03 ยท MAA Monitor, Fetal Doppler Ultrasound

Siemens Medical Solutions USA, Inc.

Acuson Diagnostic Ultrasound System Transducer; Model Number(s): L582 Linear Array Transducer for use on the 128XP Ultrasound System; Catalog Numbers: 23656 (New), 45441 (Refurb)

DateMay 22, 2003 (initiated Mar 31, 2003)
StatusTerminated
ReasonThe product has a software condition in that the surface temperature of the transducer may reach above the thermal limit.
Root causeOther
ActionOn 3/24/03, the firm issued letters via US Mail to all its direct consignees informing them of the affected device and providing instructions on the recall.
Quantity220 units
DistributionNationwide and worldwide. The product received nationwide distribution to 123 consignees: 23 Distributors (50 transducers) and consumers/users: approximately 100 customers (146 transducers). 2 transducers were consigned to Federal Government locations: (1) Martin Army Community Hospital, 9200 Marne
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K884632ACUSON I5100 ULTRASOUND TRANSDUCER
Acuson Corp., Jan 13, 1989