Recall Z-0917-03 ยท JJF Analyzer, Chemistry, Micro, For Clinical Use

Roche Molecular Systems Inc

COBAS TaqMan analyzer, Series 96; catalog number 8080216.

DateJun 5, 2003 (initiated Mar 6, 2003)
StatusTerminated
ReasonA false positive result may be reported.
Root causeOther
ActionEach customer was informed of the recall by telephone on March 6, 2003. Customers were instructed to check each postiive result to see if the 'RFITOOLOW' flag was generated, and follow-up instructions were provided. A recall letter was also faxed and mailed to each customer on March 6, 2003.
Quantity16
DistributionNationwide. California, Colorado, Georgia, Illinois, Maryland, Missouri, New Jersey, New Mexico, North Carolina, Utah and Virginia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012966ROCHE COBAS TAQMAN ANALYZER
Roche Molecular Systems, Inc., Feb 1, 2002