ClearanceBrief
Recall Z-0902-03 ยท KNC Table, Obstetric (And Accessories)

Hill-Rom, Inc.

Hill-Rom brand Affinity three birthing bed

DateMay 29, 2003 (initiated Apr 24, 2003)
StatusTerminated
ReasonPossible electrical shock hazard, as the power cord grounding pin may break off or become detached.
Root causeOther
ActionLetters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs with the flared ground pin. [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]
DistributionUnited States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915779AFFINITY BED MODEL 3600 SERIES
Hill-Rom, Inc., Feb 24, 1993