Recall Z-0869-03 ยท JQW Station, Pipetting And Diluting, For Clinical Use

Tecan U S Inc

Tecan Genesis RSP and NPS Instruments with barcode scanner type PosID2 option with firmware version 2.45 or lower.

DateMay 29, 2003 (initiated Feb 27, 2003)
StatusTerminated
ReasonFirmware defect. Under certain conditions a strip rack is no longer in the initial position after a motor overload of the barcode scanner PosID2.
Root causeOther
ActionTecan US notified consignees by registered letter on/about 3/13/2003. International accounts were notified via e-mail 3/13/2003 by Tecan Switzerland. Consignees were advised to contact the firm''s Technical Support line to coordinate the PosID firmware upgrade.
Quantity327 instruments require firmware upgrade.
DistributionInstruments requiring upgrade were distributed to various clinical laboratories and blood banks Nationwide and World wide .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953345GENESIS
Tecan U.S., Inc., Nov 21, 1995