Recall Z-0712-03 ยท JOJ Indicator, Physical/Chemical Sterilization Process

Sybron Dental Specialties

Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck 1.8%

DateApr 9, 2003 (initiated Feb 25, 2003)
StatusTerminated
ReasonMoisture ingress into packaging of test strips may cause premature failure.
Root causeOther
ActionNotice was sent to customers of the potential problem on 3/7/2003. The notice included directions to perform a Modified Quality Control Procedure. Instructions are to test strips function each day.
Quantity4,112 cases of two bottles each. 60 strips per bottle.
DistributionNationwide and Canada.