Recall Z-0789-03 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

Medtronic Ave Inc

Medtronic AVE brand BridgeAssurant Balloon Expandable Stent Delivery System for Biliary Catheter;Model #s: FB720VF, FB730 VF, FB740 VF, FB760 VF, FB820VF, FB830VF, FB840VF, FB860VF; FB720VL, FB730VL, FB820VL, FB830VL; FB720F, FB730F, FB740F, FB760F, FB820F, FB830F, FB840F, FB860F; FB720L, FB730L, FB820L, FB830L

DateMay 7, 2003 (initiated Feb 27, 2003)
StatusTerminated
ReasonDuring use, device may exhibit high insertional resistance which may result in device damage, kinking and subsequent balloon deflation difficulties.
Root causeOther
ActionA letter was sent to each US commercial accounts via UPS overnight mailing on Saturday, March 1st, 2003 for delivery on Monday, March 3rd 2003. The notification instructs firm''s customers to sequester product from the specific lot numbers so that firm''s rep could assist in removing the affected product.
DistributionProduct was released for distribution to 446 consignees nationwide. The recall was appropriately extended to the user level; i.e., Clinics, hospitals and medical Centers who received the recalled product. There were no shipments to U.S. Government facilities or foreign or Canadian distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.