Recall Z-0844-03 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

AEROSET Analyzer

DateMay 20, 2003 (initiated Aug 15, 2002)
StatusTerminated
ReasonDesign of high concentration waste line causes formation of a precipitate in the waste line causing overflow of waste.
Root causeOther
ActionThe firm sent Technical Service Bulletins dated August 15, 2002 to consignees. Field Representatives are correcting the device in the field. A customer letter was released on August 20, 2002 to provide customers actions to take to prevent the occurrance.
Quantity578
DistributionNationwide and to the following foreign countries: Mexico, Venezuela, Columbia, Argentina, Brazil, Puerto Rico, Dominican Republic, Canada, Germany, Hong Kong, New Zealand, and China