Recall Z-0684-03 ยท LCH Tonometer (Calibration And Q.C. Of Blood-Gas Instruments), Clinical

EMD Chemicals Inc

Ethanol Standard, 0.5mg Ethanol in 1 mL. Product size configuration: 1 mL glass ampule shipped 10 ampules per case. Item Number 68988/95

DateMar 26, 2003 (initiated Dec 13, 2002)
StatusTerminated
ReasonThe Ethanol Standard is OOS on the low end. The results were 0.44 mg/mL while the specification is 0.48-0.52 mg/mL.
Root causeOther
ActionAll customers were nofitied by phone and letter. The first notification occurred on 12/17/2002 followed with additional notifications on 1/24/2003, 1/28/2003, and 2/4/2003.
Quantity139 cases of 10 ampules
DistributionThe Ethanol Products were distributed nationwide by the following distributors: VWR International, 1310 Goshen Pkwy, West Chester, PA 19380; Government Scientific, 8460 K-Tyco Road, Vienna, VA 22182; Allegiance Healthcare, 1430 Waukegan Rd., McGaw Park, IL 60085; Fisher Scientific, 2000 Park Lane,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K990516ETHYL ALCOHOL CLINICAL STANDARD SOLUTIONS
Emd Chemicals, Inc., Mar 31, 1999