Recall Z-0679-03 ยท MOI System, Test, Immunological, Antigen, Tumor

Abbott Laboratories HPD/ADD

Abbott AxSYM CA 15-3 Master Calibrator Pack, list 3B42-30; the pack contains two 4-mL bottles, with Master Calibrator 1 having 0 U/mL and Master Calibrator 2 having 60 U/mL; Abbott Laboratories, Abbott Park, IL 60064

DateMar 26, 2003 (initiated Mar 4, 2003)
StatusTerminated
ReasonThe CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.
Root causeOther
ActionRecalled by letter dated 3/4/03. The accounts were informed of the depressed values for the CA 15-3 High Control and for patient specimens, and were requested to discontinue use and destroy any of the three affected lots found in inventory.
Quantity958 packs
DistributionNationwide and Ecuador, England, Hong Kong, Chile, Japan, Canada, Singapore, Australia, Pakistan, Argentina, Thailand, Aruba, Korea and Jamaica.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963926AXSYM CA 15-3
Abbott Laboratories, Sep 15, 1997